Malaysia MDA, Issuance of the 2nd edition of the guideline for the placement of HIV-Self Test (HIVST) Kit in Malaysia Market

JNM Global 제이엔엠글로벌 2025. 2. 19. 14:14

On January 31, 2025, the Medical Device Authority (MDA) of Malaysia issued the 2nd edition of the guidelines (MDA/GD/0065) to guide companies on the requirements for pre-market, placement in the market and post-market activities for HIV self-testing kits (HIVST) in the Malaysian market.

 

The key updated contents of the guidelines are as follows:

1) Regulatory requirements for facilities handling HIVST

 - Both manufacturers and Authorized Representatives (AR) must obtain establishment licenses based on their business activities and roles.

 

2) Performance requirements

- Specific performance criteria have been introduced for two types of samples as follows:

Type of test to detect antibodies and antigens/antibodies (Based on specimen type) Sensitivity Specificity
1) Using blood samples Shall not be less than 99.0 % Shall not be less than 99.0 %
2) Using saliva samples Shall not be less than 92.0 % Shall not be less than 99.0 %

 

3) Conformity assessment procedure

- Additional details on the conformity assessment process for the manufacture and import of HIVSTs, conducted by a designated Conformity Assessment Body (CAB), have been included.

 

4) Registration requirements and process flow of HIVST

- The MDA may reject application in the following cases:

 * The HIVST is wrongly classified.

 * Failure to provide additional information of the HIVST requested by the Authority within 90 days or any granted extension period.

 * Submission of the application without the requested information, particulars, or documents.

 * Non-compliance with registration requirements as evidenced by the provided documentation.

 

5) Instructions for use (IFU) requirements

 - In case of issues related to the use of the HIVST, contact information must be included so that users can file complaints.

 

6) Additional requirements on labelling

 - The HIVST label must include an audio-visual guide for the testing procedure and disposal methods, as well as a QR code linking to the TEST NOW platform for HIV result reporting and additional support.

 

Please refer to the attached document below.

FINAL GD Placement of HIV Self-test Kit in Malaysia Market 2nd Edition.pdf